Empower user cannot see area, height, or amount columns in review window when reviewing result set - WKB29887
Article number: 29887
SYMPTOMS
- Empower user cannot see area, height, or amount columns in review window when reviewing result set
- User logged in with administrator user type can view the columns in review, but a specific user type cannot
ENVIRONMENT
- Empower 3 Feature Release 1
CAUSE
User type did not have "view quantitation peak fields in review" privilege. This privilege shows the following columns in review: Area, % area, height, % height, amount, % amount, response, and concentration.
FIX or WORKAROUND
Enable the "view quantitation peak fields in review" privilege for the user type.
ADDITIONAL INFORMATION