After processing GPC standards, no calibration curve is created in Empower - WKB29342
Article number: 29342
SYMPTOMS
- After processing, no calibration curve is created
ENVIRONMENT
- Empower 3
- Gel phase chromatography (GPC)
CAUSE
Fourth-order calibration requires a minimum of five points to draw the calibration curve. The user is processing only the first injection, which contains four peaks.
FIX or WORKAROUND
Process more injections so that at least five peaks are integrated and calibrated.