Computer keeps rebooting, even while acquiring data in Empower - WKB29174
Article number: 29174
SYMPTOMS
- Computer shuts down randomly, even during analysis
ENVIRONMENT
- Empower 3 FR4
- Windows 10
CAUSE
The battery backup is set to put the computer into shutdown.
FIX or WORKAROUND
Change the battery backup setting to never shut down the computer.
ADDITIONAL INFORMATION