Instrument failure when trying to run Agilent GC - WKB286208
Article number: 286208
SYMPTOMS
- Instrument failure when trying to run Agilent GC
- Can monitor the instrument method without error
ENVIRONMENT
- Agilent GC
- Headspace
- Empower
CAUSE
The Agilent Pre-Configuration utility is configured with the Headspace
FIX or WORKAROUND
- Run the Agilent Pre-Configuration Utility and remove the headspace
- Reboot the Acquisition Node