Article number: 28418

SYMPTOMS

  • Three samples do not process during batch processing of sample set
  • Samples before and after in the sample set produce results
  • No errors in message center
  • Plenty of free tablespace

ENVIRONMENT

  • Empower 3

CAUSE

Sample set has duplicate Calibrate lines for that set of samples and no Quantitate function.

FIX or WORKAROUND

  1. Delete the second Calibrate and add a Quantitate function to the sample set.

ADDITIONAL INFORMATION

 