No results generated for a few samples in one sample set in Empower - WKB28418
Article number: 28418
SYMPTOMS
- Three samples do not process during batch processing of sample set
- Samples before and after in the sample set produce results
- No errors in message center
- Plenty of free tablespace
ENVIRONMENT
- Empower 3
CAUSE
Sample set has duplicate Calibrate lines for that set of samples and no Quantitate function.
FIX or WORKAROUND
- Delete the second Calibrate and add a Quantitate function to the sample set.