Empower installation stalls at the Building the Empower Oracle Database step - WKB282870

Article number: 282870

SYMPTOMS

  • Empower installation hangs indefinitely at the Building the Empower Oracle Database step
  • CreatePDB.BAT remains running in the task manager

clipboard_ea539935f6db53ffe8b8f8e5c28338e6f.png

clipboard_ef28a9f1a04d86785c69c2e0b41fdd8ff.png

ENVIRONMENT

  • Empower 3.6.0 or higher

CAUSE

Installation media was copied to a folder that contains spaces (e.g., F:\Media\Empower 3.8.1).

FIX or WORKAROUND

  1. Do not use spaces in the folder name or path to the media.
  2. Copy the image (e.g., 667007113DDRevA.ISO) to the server or PC.
    • Mount the image.
    • Run the setup.exe.

ADDITIONAL INFORMATION

 