Buffered Empower data is not recovered for some LAC/E32; error "Unable to create remote file because a security package specific error occurred" - WKB280001

Article number: 280001

SYMPTOMS

  • Data is buffered and will not recover for some LAC/E32
  • Test Fileservice and test share report as valid
  • Other LAC/E32 nodes upload their data correctly
  • DCOM settings confirmed correct
  • MSDTC errors in event logs
  • Issues connecting to domain controller errors in event logs
  • "Unable to create remote file because a security package specific error occurred"

clipboard_eacbced4fbabbfdf2730d74487a4bf888.png

ENVIRONMENT

  • Empower
  • System runs the WatersService

CAUSE

Authentication issues due to connectivity between the LAC/E32 and the domain controller

FIX or WORKAROUND

  1. Remove the LAC/E from the domain.
  2. Reboot.
  3. Join the LAC/E to the domain.
  4. Make test injections to the same project that was last acquired into.

ADDITIONAL INFORMATION

  • Several Windows updates were also pending and were applied when the Empower server was rebooted attempting to resolve this issue.
  • The Empower server was also the domain controller.
  • Removing and re-adding the LAC/E32 to the domain resolved the issue.