In Empower, sample name is missing when viewed as channels, injections, or results - WKB26823
Article number: 26823
SYMPTOMS
- When viewing the acquired sample set as injections, channels, or results, the sample name field does not populate
ENVIRONMENT
- Empower 3
- Project window
CAUSE
This issue is documented as PCS#56335 / CRI-372
When user altered sample sets in the Review window, saved the alterations without closing the Review window, processed the data, and then saved the results, the Sample Set ID, Sample Name, or other sample and sample set fields were missing from the results.
FIX or WORKAROUND
The issue is resolved in Empower 3.6.1
ADDITIONAL INFORMATION