PDA detector error "Data Buffer Overflow" - WKB267745

Article number: 267745

SYMPTOMS

  • A run starts and the user returns later to see the PDA detector error "Data Buffer Overflow"

ENVIRONMENT

  • Empower
  • PDA detector

CAUSE

The acquisition node sleep setting was not disabled, and the node went to sleep during the run.

FIX or WORKAROUND

Configure the power and sleep settings as described in the installation and configuration guide.

ADDITIONAL INFORMATION

What are the Empower 3 installation guidelines and minimum hardware requirements? - WKB61110