Empower preferences not saved in Citrix session - WKB26714
Article number: 26714
SYMPTOMS
- Empower preferences are not saved if the Run Samples window is not maximized while in a Citrix session.
ENVIRONMENT
- Empower 3 FR3 HF1
- Citrix XenApp v7.6
CAUSE
Possible issue with Citrix Seamless settings.
FIX or WORKAROUND
Altering the Windows registry corrected this issue in several cases.
Execute the following steps on all Citrix servers supporting Empower.
- Navigate to the registry key HKEY_LOCAL_MACHINE/System/CurrentControlSet/Control/Citrix/wfshell/TWI.
- Create a Value Type: REG_DWORD.
- Create a Value Name: SeamlessFlags.
- Set the SeamlessFlag value to 0x26DEA7.
ADDITIONAL INFORMATION
See this link for further detail.
Discuss this change with GSS before implementation.
Changing the registry can cause significant damage to the operating system. Back up the registry before implementing this change.