Processing issue with ICF Support Layer 3.7.0 in Empower - WKB267176
Article number: 267176
SYMPTOMS
- When using ICF Support Layer 3.7.0 with Empower (any version) and processing data that was acquired on an Agilent system that is controlled by this driver, results cannot be created
- This problem occurs whether processing is done using a newly created processing method (or method set) or using a processing method (or method set) that was created using a previous version of ICF Support Layer
- When this issue occurs, a result set is present, but there are no results within the result set
- Error messages of the Processing category stating “There are no channels specified to clear in method set [method set name]” and “This is usually caused by not specifying an instrument method in the method set – Processing aborted” appear in the Empower Message Center
ENVIRONMENT
- Agilent ICF SL 3.7.0
- Empower
- Agilent LC
- Agilent GC
CAUSE
N/A
FIX or WORKAROUND
No workaround exists for this issue except to uninstall ICFSL 3.7.0 and reinstall an earlier version.
ADDITIONAL INFORMATION
CRI-6148 published on Waters website
https://www.waters.com/waters/suppor...6148&type=DFCT