The channel name from the method set is different than the instrument method using Beckman CE , LIF detector, and Empower - WKB26689
Article number: 26689
SYMPTOMS
- When Voltage Current LIF Channels 1 and 2 are collected, you can see the channels and review the chromatograms, but when you derive a channel in Method Set you cannot derive LIF channel or select it from the table
- Can see only PDA 254 and PDA 214, although the method does not monitor these channels
- The LIF channels are not available to select from or derive, so when processing, unwanted results, such as Current, Voltage, and LIF Channel 2, are produced
ENVIRONMENT
- Empower 3 Feature Release 3 Hotfix 1
- Beckman Coulter PACE MDQ 1.2.1.0 driver with LIF detector
CAUSE
Defect PCS-56179: "Method set shows different channels than instrument method"
FIX or WORKAROUND
Workaround
- Manually type the channel name in the method set editor.
- The drop-down list also allows typing the channel name.
ADDITIONAL INFORMATION
id26689, EMP2LIC, EMP2OPT, EMP2SW, EMP3GC, EMP3LIC, EMP3OPT, EMP3SW, EMPGC, EMPGPC, EMPLIC, EMPOWER2, EMPOWER3, EMPSW, SUP