New projects do not have the expected project-level Comment or Confirm Identity settings in Empower 3.8.0 - WKB265525
Article number: 265525
SYMPTOMS
- Newly created projects do not have the expected Comment and Confirm Identify settings
- New projects are instead set to Comment = Silent and Confirm Identity = False
ENVIRONMENT
- Empower 3.8.0
CAUSE
This behavior is documented as CRI-6115 against Empower 3.8.0.
FIX or WORKAROUND
- Disable the setting "Don't allow user to change default 'Require User Comments On' settings" on the New Project tab in System Policies and implement a procedural control on new project creation, advising users not to change the new project default settings.
- Clone from a project known to have the expected project-level Comment and Confirm Identity settings, because cloning is not affected by this behavior.
- CRI-6115 is resolved in Empower 3 Version 3.8.0.1
ADDITIONAL INFORMATION
CRI-6115 is resolved in Empower 3 Version 3.8.0.1. Please refer the release notes 715008910 Ver. 00 Page No: 20 in this regard.