Sample set interrupted during Windows logoff - WKB262252

Article number: 262252

SYMPTOMS

  • When sample set is running and user logs off from Windows, the sample set is interrupted without any error messages
  • When system is in monitoring mode and user logs off from Windows, the system goes to idle state

ENVIRONMENT

  • Empower (all versions)

CAUSE

When the user logs off from Windows, Empower-related services running in the background are killed by the operating system, so the sample set is interrupted.

FIX or WORKAROUND

Reload the operating system.

ADDITIONAL INFORMATION

Not able to find a solution? Click here to request help.