When I manually delete skimmed peaks, the results for area are inaccurate in Empower 3 - WKB2617

Article number: 2617

SYMPTOMS

  • When deleting multiple peaks from a chromatogram, if one or more of the peaks are tangential or exponential skimmed peaks, the area of the skimmed peaks are not added correctly to the main peak area.

PCS 54843_1.PNG

PCS 54843_2.PNG

PCS 54843_3.PNG

ENVIRONMENT

  • Empower 3

CAUSE

This is a defect reported under PCS 54843.

FIX or WORKAROUND

This issue is resolved with the Empower 3 Service Release 2 Hotfix 3.

ADDITIONAL INFORMATION