SYMPTOMS

  • When deleting multiple peaks from a chromatogram, if one or more of the peaks are tangential or exponential skimmed peaks, the area of the skimmed peaks are not added correctly to the main peak area.

ENVIRONMENT

  • Empower 3

CAUSE

This is a defect reported under PCS 54843.

FIX or WORKAROUND

This issue is resolved with the Empower 3 Service Release 2 Hotfix 3.

ADDITIONAL INFORMATION