Skip to main content
Waters

ORA-12528 Empower login error - WKB252276

Article number: 252276

SYMPTOMS

  • ORA-12528 Empower login error

ENVIRONMENT

  • Empower (all platforms)

CAUSE

The Windows KB5022289 patch was applied prior to the error.

FIX or WORKAROUND

Go to Services as a Windows administrator and restart the Oracle Empower TNS Listener and then the Oracle Service Empower.

ADDITIONAL INFORMATION

 

Not able to find a solution? Click here to request help.