With Empower 3, a communication error appears on the ACQUITY Arc Direct Control Panel, and an instrument failure occurs - WKB250563
Article number: 250563
SYMPTOMS
- A communication error appears on the Direct Control Panel for QSM-R
- Right-clicking on the QSM-R Direct Control Panel does not clear the communication error
ENVIRONMENT
- Empower 3 Personal
- ACQUITY Arc
- ACQUITY UPLC H-Class
CAUSE
Communications faults
Dead instrument battery
FIX or WORKAROUND
- Check the video at the following URL and perform Stop Acquity: https:/videos.waters.com/ja/detail/videos/%E3%82%B5%E3%83%BC%E3%83%93%E3%82%B9/video/6015500414001?autoStart=true
(This video is available in Japanese only.) - If this does not help, see the video at the following URL and check if the instrument is recognized.
If the recognition of QSM-R without serial is Yes, contact Waters because the battery may be dead:
https://videos.waters.com/ja/detail/videos/%E3%82%B5%E3%83%BC%E3%83%93%E3%82%B9/video/6021474158001?autoStart=true
(This video is available in Japanese only.) - If the problem persists, try shutting down and restarting the instrument and PC, if you have not already done so.
The order to power-on is turn on the PC → start the PC OS → power-on the instrument → launch Empower after the initialization diagnosis of the instrument is completed.
For LAC/E-connected instruments, the LAC/E must be restarted. - If no improvement, check the video at the following URL and see if there is a response from the QSM-R address:
https:/videos.waters.com/ja/detail/videos/%E3%82%B5%E3%83%BC%E3%83%93%E3%82%B9/video/6021470052001?autoStart=true (This video is available in Japanese only.)