Empower Run Samples error: Database error, unable to continue - WKB2502
Article number: 2502
SYMPTOMS
- Error message "Database error, unable to continue" appears when trying to run a previously run Sample Set method.
ENVIRONMENT
- Empower 2
- Empower 3
CAUSE
Corrupt Sample Set method
FIX or WORKAROUND
- Create a new Sample Set method from scratch.
ADDITIONAL INFORMATION
Confirm that the project has enough free tablespace (at least 10 MB) to save the new Sample Set method. If not , increase the project quota to free up tablespace.