Failure to get ready to run an acquisition with Agilent LC - WKB250090
Article number: 250090
SYMPTOMS
- Aborted by another module error in message center and diagnostics logs
- Failed to get ready to run an acquisition in ICF logs.
- Condition column or Equilibration line executes in sample set without any error, but instrument failure appears while executing sample injection
ENVIRONMENT
- Empower 3.6.1
- Empower version 3.6.1.1
- ICF 3.6
- ICF 3.5.1
- ICF version 3.6.1.3
- Agilent LC
- Agilent GC
CAUSE
A post run time of 1 min was added in the instrument method, creating this issue.
FIX or WORKAROUND
Remove the post run time that was added in the instrument method.
or
Non-inject functions with PostRun Time should be run separately in a different sample set. refer the Defect ID INFICFSL-1973 with ICF Support Layer 3.6.1.
ADDITIONAL INFORMATION