How do you manually push buffered Empower data to the raw data share on the Empower server? - WKB249105
Article number: 249105
ENVIRONMENT
- Empower
- LAC/E32
ANSWER
Run the command InstrumentServer - recover from a command prompt at c:\empower\bin on the desktop of the LAC/E32. You can also reboot the LAC/E32 as long as no acquisition is running at the time. Consult the LAC/E32 recover.log or view the data in Empower's Review window to confirm that the data uploaded to the server. You may need to wait ten minutes after the reboot to see the data recovered and viewable.