Skip to main content
Waters

Amounts calculate as 0.00 with Empower 3 - WKB249069

Article number: 249069

SYMPTOMS

  • Peaks integrate correctly
  • Injected one standard five times at the same volume to generate standard curve

ENVIRONMENT

  • Empower 3
  • Agilent GC

CAUSE

There is a value on the Purity tab of Amounts.

FIX or WORKAROUND

Purity is not in use, so remove this value to obtain appropriate results. 

ADDITIONAL INFORMATION

 

Not able to find a solution? Click here to request help.