All non-administrator users in Empower can run samples but cannot process and create results - WKB247972
Article number: 247972
SYMPTOMS
- All users except administrators can run samples but cannot process and create results
- Process button is dimmed (unavailable)
ENVIRONMENT
- Empower 3.7.0
CAUSE
Users do not have the following privileges:
- Save Results
- Save Results and Calibrations in review
FIX or WORKAROUND
- From the main Empower window, go to Configure the System.
- Highlight "user type" on the left-hand side of the window.
- Highlight the user's user type on the right-hand side of the window.
- Right-click on the user's user type and choose "properties".
- Click the "Methods" tab.
- Scroll down and find the privilege Save Results and "Save results and calibrations in review". Select these privileges as shown in screenshot below.