ACQUITY communications lost, console dimmed - WKB241233

Article number: 241233

SYMPTOMS

  • No communications to ACQUITY; dimmed windows in Run Samples and dimmed and missing info in the Console
  • Limited control over SM; can prime syringes but no live diagnostics; everything else unresponsive
  • Antivirus and other issues ruled out; instruments can be pinged; other (Alliance) insturments on LAC/E communicate OK

ENVIRONMENT

  • Empower
  • ACQUITY
  • LAC/E

CAUSE

Unknown        

FIX or WORKAROUND

  1. Uninstall and reinstall the ICS on the LAC/E (communications restored and stable).
  2. Reinstall only ICs required for installed modules (previously, all ICs were on the LAC/E).