ACQUITY communications lost, console dimmed - WKB241233
Article number: 241233
SYMPTOMS
- No communications to ACQUITY; dimmed windows in Run Samples and dimmed and missing info in the Console
- Limited control over SM; can prime syringes but no live diagnostics; everything else unresponsive
- Antivirus and other issues ruled out; instruments can be pinged; other (Alliance) insturments on LAC/E communicate OK
ENVIRONMENT
- Empower
- ACQUITY
- LAC/E
CAUSE
Unknown
FIX or WORKAROUND
- Uninstall and reinstall the ICS on the LAC/E (communications restored and stable).
- Reinstall only ICs required for installed modules (previously, all ICs were on the LAC/E).
ADDITIONAL INFORMATION
id241233, comms