Error: Cannot continue background processing a sample set because sample set XXXX has one or more channels that are still being acquired - WKB240773
Article number: 240773
SYMPTOMS
- Unable to process the sample set
- The following error message appears in the message center while processing the sample set:
"Cannot continue background processing a sample set because sample set XXXX has one or more channels that are still being acquired."
ENVIRONMENT
- Empower 3
CAUSE
DCOM settings are incorrect.
Microsoft hardening patches are installed
FIX or WORKAROUND
Apply the DCOM settings as described in the installation and configuration guide.
Refer to KCS article WKB230033 and the registry fix in KB5004442.