No Mw, Mn, or Mp values for GPC broad unknown peaks in Empower after processing - WKB23845
Article number: 23845
SYMPTOMS
- After processing, no Mw, Mn, or Mp values for GPC broad unknown peaks in Empower
- Valid curve created from three narrow standards
- Sample has sample type correct as Broad Unknown in Alter Sample
ENVIRONMENT
- Empower
- GPC
CAUSE
The peaks were not listed in the slicing tab. Therefore, no values for the broad unknown are calculated.
FIX or WORKAROUND
- Open the result set in Review.
- Verify that the correct peaks are integrated.
- Open the processing method window and go to the slicing tab.
- Go to Options > Fill from Results. The slicing tab populates with peaks of interest.
- Re-quantitate the broad unknown.
ADDITIONAL INFORMATION