FTN temperature setting fault - WKB236973

Article number: 236973

SYMPTOMS

  • FTN temperature setting fault. Entered value is multiplied by 10
  • No error messages
  • Regional settings correct
  • Other temperature settings not affected

ENVIRONMENT

  • ACQUITY UPLC I-Class SM-FTN
  • ACQUITY UPLC H-Class SM-FTN
  • ACQUITY UPLC H-Class Bio SM-FTN
  • Empower Enterprise 

CAUSE

Corrupt driver installation on LAC/E

FIX or WORKAROUND

Reinstall the driver pack on the LAC/E.

ADDITIONAL INFORMATION