Why is the pass/fail criterion for wavelength accuracy qualification on the fluorescence detector ±6 nm? - WKB236967
Article number: 236967
ENVIRONMENT
- Empower
- System SQT
ANSWER
- To start with, it is important to note that the light in a fluorescence detector passes through two monochrometers before being detected.
- Wavelength accuracy of the Waters ACQUITY FLR emission monochrometer is ±2 nm.
- Wavelength accuracy of the Waters ACQUITY FLR excitation monochrometer is ±2 nm.
- This means that the tolerance with respect to total wavelength accuracy is increased by at least the square root of 2 times the tolerance of each individual monochrometer, so this value is rounded to a combined specification of ±3 nm.
- For the qualification tests employed by the SystemsQT, a second-order reflection is used, so the wavelength tolerance of the second-order output must be doubled to ±6 nm.
- This is why the qualification pass/fail criterion is listed as ±6 nm.
ADDITIONAL INFORMATION
id236967, 2475, 2475B, A-10FL, A-30FL, EMP2LIC, EMP2OPT, EMP2QLIC, EMP2SW, EMP3GC, EMP3LIC, EMP3OPT, EMP3Q1ULIC, EMP3QLIC, EMP3SW, EMPGC, EMPGPC, EMPLIC, EMPOWER2, EMPOWER3, EMPSW, SUP, UPBINARY, UPFLRARC, UPFLRDET, UPPFLR