Are run buffer and ring buffer the same with Agilent and Empower? - WKB23650

Article number: 23650

ENVIRONMENT

  • Empower 2 Feature Release 5
  • Agilent 1.06 driver

ANSWER

Ring buffer and run buffer are considered the same from an Empower perspective, which means that the detector's buffer/memory is full because it can't offload the data. When that happens, it sends a message to Empower saying that it is full, and Empower stops the run. The detector is not able to process/move the data as quickly as it is being acquired.

ADDITIONAL INFORMATION

Errors reported are:

  • Extract Data: Software has failed to append raw data to the instrument ring buffer. 
  • G1314A: Signal run buffer full. Storing of signal data stopped.