Empower message center error: Instrument failure - End of run wait timed out at xxxxx seconds. No reply from W410 XXXXXXXXXX - WKB233808
Article number: 233808
SYMPTOMS
- Empower message center error "Instrument failure - End of run wait timed out at xxxxx seconds. No reply from W410# XXXXXXXXXX"
- Error occurs after first injection is complete
- Node properties shows status = Yes for all modules
- WDHCP looks good
ENVIRONMENT
- Empower 3 Feature Release 5 Service Release 4 Enterprise
- Alliance HPLC System
- 2695 Separations Module
- 2695D Dissolution Separations Module
- 2489 UV/Vis Detector
- 410/2414 Refractive Index Detector
CAUSE
- System configuration issue
- The 410 is OFF in the instrument method
FIX or WORKAROUND
- Remove the module that is not part of the instrument method. In this example, the 2414 is not being used.
- Create a separate system with just the modules that are part of the method.
- Save the method.
- Run samples.
ADDITIONAL INFORMATION
id233808, 2414, 2414B, 2489, 2489B, 2489SFC, 2690, 2690D, 2690DUPGD, 2690UP, 2695, 2695D, 2695DE, 2695DEB, 2695DUP, 2695E, 2695EB, 2695INERT, 2695UP, A-10RI, A-10SSM, A-10UV