Error message "Instrument failure - Method setup timed out after 120.1 seconds. No reply from W1500-CH#xxxxxxxxxx" - WKB233316

Article number: 233316

SYMPTOMS

  • "Instrument failure - Method setup timed out after 120.1 seconds. No reply from W1500-CH#xxxxxxxxxx"
  • Unable to run samples - Instrument failure
  • All readbacks in Empower QS are at set point 
  • Column Heater shows last set point of 35 °C in Empower QS

ENVIRONMENT

  • Empower QS
  • Modular LC (1525 Isocratic Pump, 7725i Manual Injector, Column Heater, 2489 UV detector)
  • Empower 3
  • Waters UPLC
  • Waters detectors
  • Waters HPLC
     

CAUSE

The column heater power switch was accidentally turned off. 

FIX or WORKAROUND

Power-on the Column Heater module by moving the switch on the side from O to I.

ADDITIONAL INFORMATION

If all instruments are on and you still get the error, verify the instrument method or recreate the instrument method. 