Beginning with Empower 3 FR5, timestamps may display timezone offset instead of the timezone abbreviation - WKB228794
Article number: 228794
SYMPTOMS
- Empower timestamps may show a timezone offset instead of the expected timezone abbreviation
- Prior to Empower 3 FR5, the timestamp displays as follows:

- Beginning with Empower 3 FR5, the timestamp displays as follows:

ENVIRONMENT
- Empower 3 FR5 and later
- Empower 3.6.x and later
CAUSE
Changes made by Oracle in the Oracle timezone files have removed various timezone abbreviations. As a result, Empower displays the timezone offset instead of the timezone abbreviation for affected timezones.
FIX or WORKAROUND
Set Empower System Policies to "Show Region Name" instead of "Show Region Abbreviation".
ADDITIONAL INFORMATION