One project backup is always missing the EXP and INF files - WKB228752
Article number: 228752
SYMPTOMS
- One project backup in particular is always missing the EXP and INF files
- Other backups are correct and can be restored
ENVIRONMENT
- Empower
CAUSE
Disk (HDD) corruption on the raw data share. The backup process could not read some raw data files, so the backup could not complete correctly.
FIX or WORKAROUND
- Repair the disk errors.
- Locate a backup of the project prior to the disk corruption.