With ICF SL 3.5.1, Agilent GC does not make injections; "Method Setup Failed" in message center - WKB227033
Article number: 227033
SYMPTOMS
- Agilent GC does not make injections
- "Method Setup Failed" appears in message center
ENVIRONMENT
- Empower 3
- ICF SL 3.5.1
CAUSE
"Prep Run..." is selected.

FIX or WORKAROUND
Clear the selection for ALS/HS.
ADDITIONAL INFORMATION
In previous ICF, this was set in Agilent Preconfiguration.
Methods from the previous ICF may have this selected.