Empower Configuration Manager fails when multiple projects are backed up - WKB226963

Article number: 226963

SYMPTOMS

  • Empower Configuration Manager fails when multiple projects are backed up in a single action
  • The failure appears immediately after starting the backup
  • Individual projects can be backed up successfully

ENVIRONMENT

  • Empower 3 Feature Release 5 Service Release 4

CAUSE

The permissions on the directory x:\empower\logs does not allow the OS user to write files in this location. This is the default location where the multiple-project backup or restore creates summary log files.  

FIX or WORKAROUND

Configure the permissions at the OS level to allow the users running Empower to write to this directory.

ADDITIONAL INFORMATION

 