Skip to main content
Waters

Error with Empower 3 and Agilent 7890 GC: "Method does not match current hardware configuration and requires manual resolution" - WKB226167

Article number: 226167

SYMPTOMS

  • Recorded in Message Center: "Method does not match current hardware configuration and requires manual resolution"
  • Even if set up, the numbers according to the instrument method are not reflected on the main unit
  • Last time used without 7697, presets were not reset; specified GC Back in the instrument method option while the 7697 was still connected
  • This time using headspace
  • If Yes selected for "Do you want to change to the current instrument configuration when you open the instrument method?", issue does not improve
  • Rebooting LAC/E and resetting presets does not improve

ENVIRONMENT

  • Agilent 7890 GC
  • Agilent 7697A Headspace

CAUSE

 

FIX or WORKAROUND

  1. Restart the LAC/E with the Empower screen closed.
  2. Create a new instrument method.

ADDITIONAL INFORMATION

 

id226167, 7890 GC, 7890 GCDET, 7890BGC, EMP2LIC, EMP2OPT, EMP2SW, EMP3GC, EMP3LIC, EMP3OPT, EMP3SW, EMPGC, EMPGPC, EMPLIC, EMPOWER2, EMPOWER3, EMPSW, HS7697B, SUP

Not able to find a solution? Click here to request help.