"Unable to locate the processing server" error when processing a sample set in Empower - WKB22601
Article number: 22601
SYMPTOMS
- "Unable to locate the processing server error, ProcMon exiting" error when processing a sample set in Empower
- Error occurs when processing one sample set in the project; all other sample sets process correctly
- All users experience the same error
ENVIRONMENT
- Empower 3 Enterprise
CAUSE
Corrupt processing method
FIX or WORKAROUND
Create a new processing method and reprocess the sample set.