Error (when copying report methods in Empower): Failed to store archived object to the database - WKB2234
Article number: 2234
SYMPTOMS
- Error message appears when copying report methods from one project to another in Empower
- "Failed to store archived object to the database"
ENVIRONMENT
- Empower
CAUSE
The destination/target project is low (less than 10 MB free) on free project tablespace.
FIX or WORKAROUND
- Increase the destination project's free tablespace.
ADDITIONAL INFORMATION