OBJECTIVE or GOAL

  • To split a shoulder peak with a tangent

ENVIRONMENT

  • Empower 3
  • Empower 2

PROCEDURE

About skim
There are two types of skim events:
Tangential skims: Draw a line (tangent line) to skim the unresolved peak from the larger peak.
Exponential skims: Draw a curve (exponential function) to skim the unresolved peak from the larger peak.
Skim method
1. Check the height ratio of each peak to determine to split with a drop line or perform the skim.
If ratio > value... Skim is performed
If ratio < value... A drop line is drawn

Value: [Value] of the skim event row

 

clipboard_e5f7a8aa31d01a7c68e1d73e399bc35ca.png

clipboard_edd14fe1e6668c31e2b0c3cdde33291d8.png

Select "Processing Method" from the window to open the Edit Processing Method window.
Click the Integration tab and add a row of Tangent skim to the event table.

clipboard_e9309629879da6ff540ccea45311ad7cc.png

Check the integration on the main window.

clipboard_e85501792b500ce9da2c99481ce41396a.png

After confirming the integration, save the skim event settings set in Save > Processing Method from the File menu.

ADDITIONAL INFORMATION

 