Error: Chrom open for read failed, file \\SERVERNAME\Waters Projects$\ProjectName\Datafile.dat, err = 'The user name or password is incorrect' - WKB221619
Article number: 221619
SYMPTOMS
- Result set generated with NO peak labels
- Error message in Empower message center: "Chrom open for read failed, file \\SERVERNAME\Waters Projects$\ProjectName\Datafile.dat, err = 'The user name or password is incorrect'"
ENVIRONMENT
- Empower 3 Enterprise
CAUSE
- Lack of permission to the raw data share path for the user
- User changed the Windows login password during the Empower session
FIX or WORKAROUND
- Ensure that the user can access the raw data share path with their credentials.
- The Windows login password should not be changed when an Empower session (including Citrix session) is open or idle.
- Log off and log in again from Windows in all active clients whenever a user password is changed or locked.
- Recommendation: Do not wait until the password expires. Change the Windows password before the expiry.
ADDITIONAL INFORMATION