In Empower, PDA spectrum shape is unusual - WKB221302
Article number: 221302
SYMPTOMS
- The shape of the PDA spectrum displayed during 3D data integration is different from the usual one, and it swings significantly negatively
- If you check the spectrum of the peak top time in the spectrum extraction box, the shape is as usual
- If you clear the "Baseline correction" check box on the spectrum table, the spectrum is as usual
ENVIRONMENT
- Empower 3
CAUSE
Immediately behind the target, there was elution of a component with large absorption, and the distance to the target was closer than usual.
Therefore, baseline correction had a strong effect.
FIX or WORKAROUND
Change the method to improve peak separation by reducing dead volumes or by changing columns or analysis conditions.