Empower installation stalls at the Building the Empower Oracle Database step - WKB219602

Article number: 219602

SYMPTOMS

  • Empower installation progress stalls at the Building the Empower Oracle Database step

clipboard_e4f198c604aad9fdfe97756844aee5add.png

ENVIRONMENT

  • Empower Enterprise
  • Empower Workgroup
  • Empower Personal

CAUSE

The Operating System user used to perform Empower installation does not have complete Administrative privileges on the computer.

FIX or WORKAROUND

  1. Ensure the Operating System user used to perform Empower installation is a member of the local Administrators group on the computer. This can be verified in Computer Management > Local Users and Groups > Groups > Administrators

clipboard_e6630bf5bf6b7529cba3862cefaa41894.png

 2. Ensure the Operation System user used to perform Empower installation is assigned the Administrator user type in Control Panel > User Accounts > Manage User Accounts. Select the user of interest and select Properties.

clipboard_e2026be75d52b832418eda609aa9f2314.png

ADDITIONAL INFORMATION

 