When using an ACQUITY FLR 3D instrument method, the derived channel is missing from the method set - WKB219456

Article number: 219456

SYMPTOMS

  • 3D data collected from an ACQUITY FLR
  • In the Derived Channel window, within the Method Set, there is no option for a channel for the FLR in the drop-down menu

clipboard_e89823134ca1154974f726eba91973d26.png

ENVIRONMENT

  • ACQUITY UPLC FLR Detector
  • Empower 3

CAUSE

There is no extracted channel saved in the method set.

FIX or WORKAROUND

  1. Bring the Sample Set with the 3D FLR data into review.
  2. Select "View" > "3D Layout".
  3. Select "Process" > "Extract Chromatogram".
  4. Use the slide rule to select the correct Wavelength. 
  5. Select "Window" > "Method Set...".
  6. The derived channel appears in the Channel Name column.
  7. Make any integration changes to the Processing Method.
  8. Select "File" > "Save" > "All".