In Empower, the Agilent Setup and Monitor button freezes - WKB21851
Article number: 21851
SYMPTOMS
- The Setup and Monitor button "freezes" when selected
ENVIRONMENT
- Empower
- Agilent
- Run Sample window
CAUSE
When the instrument freezes, there is a configuration difference between the method in use and the system configured.
FIX or WORKAROUND
Before selecting Setup and Monitor in the Run Sample window, select Edit.
A message appears if the instrument method does not match the instrument configuration based on the Preconfiguration utility and the actual instrument. The user should select "yes" when asked whether to apply a configuration change. The instrument method adapts to the current configuration and, once saved and applied as setup and monitor, it will not freeze.
ADDITIONAL INFORMATION