CE analysis is missing percent time corrected area and time corrected area - WKB2133

Article number: 2133

SYMPTOMS

  • Percent (%) time corr. (corrected) area and time corr. (corrected) area are not reported in the Review window when processing CE results
  • Percent (%) time corr. (corrected) area and time corr. (corrected) area are not available under table properties

ENVIRONMENT

  • Empower 3
  • CE

CAUSE

The processing method does not have Processing type set to CE.

FIX or WORKAROUND

  1. Create a new processing method by choosing CE from the "Processing Type".
  2. Save the method and open the method to process data. (To confirm it's a CE method, "CE" should appear in the far-right corner of the Processing window.)

See attached screen shots.

CE Processing Method.PNG

CE Method.PNG

 

CE Table Properties.PNG

ADDITIONAL INFORMATION

Ensure that the CE option is enabled under Project Properties if you want to run a CE analysis.

Click Configure System in Empower > Projects. Highlight the project name, right click, select Properties > Enabled options, and click CE/CIA.

This option comes with Empower; no separate purchase is required.

CE option.PNG