CE analysis is missing percent time corrected area and time corrected area - WKB2133
Article number: 2133
SYMPTOMS
- Percent (%) time corr. (corrected) area and time corr. (corrected) area are not reported in the Review window when processing CE results
- Percent (%) time corr. (corrected) area and time corr. (corrected) area are not available under table properties
ENVIRONMENT
- Empower 3
- CE
CAUSE
The processing method does not have Processing type set to CE.
FIX or WORKAROUND
- Create a new processing method by choosing CE from the "Processing Type".
- Save the method and open the method to process data. (To confirm it's a CE method, "CE" should appear in the far-right corner of the Processing window.)
See attached screen shots.
ADDITIONAL INFORMATION
Ensure that the CE option is enabled under Project Properties if you want to run a CE analysis.
Click Configure System in Empower > Projects. Highlight the project name, right click, select Properties > Enabled options, and click CE/CIA.
This option comes with Empower; no separate purchase is required.