Shimadzu GC collects data only for 0.01 minute with error "The GC program has been started with data acquisition cannot performed" in the Empower message center - WKB206045

Article number: 206045

SYMPTOMS

  • The injections on a sample set have acquired for only 0.01 minutes on Shimadzu GC
  • Message center records error "The GC program has been started with data acquisition cannot performed"
  • The issue occurs while using Headspace sampler

ENVIRONMENT

  • Empower 3
  • Shimadzu GC Driver 3.1
  • Shimadzu GC-2030
  • Shimadzu HS-20

CAUSE

Reconfigured from the configuration wizard before creating a new method

FIX or WORKAROUND

The issue is fixed in Shimadzu GC Driver 3.2.

ADDITIONAL INFORMATION

 