Error reported after running samples for a long time: Insufficient memory, or the database connection is terminated ORA-03135 - WKB205948
Article number: 205948
SYMPTOMS
-
After running the sample for a long time, an error is reported: Insufficient memory, or the database connection is terminated ORA-03135 error
-
The software loses the connection; you need to restart the computer again to reconnect
ENVIRONMENT
- Windows 10 LTSB 1607
- Empower 3
- Alliance HPLC System
CAUSE
When the physical memory is low (for example, only 16G) due to lengthy sampling, the PDA has a large amount of data, which causes the virtual memory of the operating system to overflow. The database program is actively terminated by the system, causing errors.
FIX or WORKAROUND
- Increase the physical memory of the computer, or set the maximum and minimum operating system virtual memory to four times the physical memory capacity.
- To release memory, restart the computer when the system is idle.