Empower - Data buffering on all LAC/E - WKB203101

Article number: 203101

SYMPTOMS

  • Raw data are not sent to the Empower server
  • No new channels in Empower project
  • Raw Data Share is valid
  • Failed to recover datafile in the Empower message center
  • Manual recover is working (instrumentserver -recover)

ENVIRONMENT

  • Empower 3 Enterprise

CAUSE

On the Empower raw data server, user groups are missing from the Windows "Users" group:
NT AUTHORITY\Authenticated Users
NT AUTHORITY\INTERACTIVE

FIX or WORKAROUND

Add again the NT AUTHORITY\Authenticated Users and NT AUTHORITY\INTERACTIVE members to the Windows user group "Users" of the Empower raw data server.