Skip to main content
Waters

Configuration mismatch error for Agilent ELSD in Empower Run Samples window - WKB202068

Article number: 202068

SYMPTOMS

  • Configuration mismatch error in Empower Run Samples window for Agilent ELSD

ENVIRONMENT

  • Empower 3 Enterprise
  • Agilent Instrument Control Framework
  • Agilent 7102A ELSD

 

CAUSE

Agilent ELSD not configured correctly on Agilent Preconfiguration Utility

FIX or WORKAROUND

  1. Check the "Cooled options" in the Agilent configuration utility for the ELSD detector. Ensure that it matches the hardware configuration. 
  2. Most of the time the error will be resolved by clearing the "Cooled Options" selection.

ADDITIONAL INFORMATION

 

Not able to find a solution? Click here to request help.