Agilent GC takes injection volume from the instrument method instead of the sample set - WKB199588
Article number: 199588
SYMPTOMS
- Agilent GC takes injection volume from the instrument method instead of the sample set
ENVIRONMENT
- Empower 3
- Agilent Instrument Control Framework
- Agilent 6890 GC
- Agilent 7890 GC
CAUSE
The injection type in the instrument method is selected as “Multiple Injections".
FIX or WORKAROUND
Change the injection type in ALS > Injector in instrument method from “Multiple Injections” to “Standard".
ADDITIONAL INFORMATION
When injection type is selected as “Multiple Injections in instrument method”, the instrument method overrides the injection volume mentioned in the sample set method, which means the autosampler will use the injection volume specified in the instrument method.
In cases where “Standard” injection type is selected, the sample set method overrides the injection volume specified in the instrument method and the autosampler uses the injection volume noted in the sample set method.