Empower 3 cannot restore child projects that have a period in their name - WKB19885
Article number: 19885
SYMPTOMS
- When restoring a hierarchy of projects, child projects with a period (".") in their name are not restored
- Child projects that don't have a period in their name restore successfully
ENVIRONMENT
- Empower 3
CAUSE
This is a defect in the Empower software.
FIX or WORKAROUND
- Restore the child projects that have a period in their name individually. (They can be restored within an existing hierarchy.)
- Avoid the use of periods in new or restored project names.